Bone marrow dosimetry in low volume mHSPC patients receiving Lu-177-PSMA therapy using SPECT/CT

Grob, D., Privé, B.M., Muselaers, C.H.J. et al. Bone marrow dosimetry in low volume mHSPC patients receiving Lu-177-PSMA therapy using SPECT/CT. EJNMMI Phys 11, 34 (2024).

https://doi.org/10.1186/s40658-024-00636-0

Abstract

Background

Bone marrow toxicity in advanced prostate cancer patients who receive [177Lu]Lu-PSMA-617 is a well-known concern. In early stage patients; e.g. low volume metastatic hormone sensitive prostate cancer (mHSPC) patients, prevention of late bone marrow toxicity is even more crucial due to longer life expectancy. To date, bone marrow dosimetry is primarily performed using blood sampling. This method is time consuming and does not account for possible active bone marrow uptake. Therefore other methodologies are investigated. We calculated the bone marrow absorbed dose for [177Lu]Lu-PSMA-617 in mHSPC patients using SPECT/CT imaging and compared it to the blood sampling method as reference.

Methods

Eight mHSPC patients underwent two cycles (3 and 6 GBq) of [177Lu]Lu-PSMA-617 therapy. After each cycle, five time point (1 h, 1 day, 2 days, 3 days, 7 days) SPECT/CT was performed at kidney level. Bone marrow dosimetry was performed using commercial software by drawing ten 1.5 cm diameter spheres in the lowest ten vertebrae to determine the time-integrated activity. Simplified protocols using only 2 imaging time points and 3 vertebrae were also compared. Blood-based dosimetry was based on the blood sampling method according to the EANM guideline.

Results

Mean bone marrow absorbed dose was significantly different (p < 0.01) for the imaging based method (25.4 ± 8.7 mGy/GBq) and the blood based method (17.2 ± 3.4 mGy/GBq), with an increasing absorbed dose ratio between both methods over time. Bland Altman analysis of both simplification steps showed that differences in absorbed dose were all within the 95% limits of agreement.

Conclusion

This study showed that bone marrow absorbed dose after [177Lu]Lu-PSMA-617 can be determined using an imaging-based method of the lower vertebrae, and simplified using 2 time points (1 and 7 days) and 3 vertebrae. An increasing absorbed dose ratio over time between the imaging-based method and blood-based method suggests that there might be specific bone marrow binding of [177Lu]Lu-PSMA-617.

Additional information

This article is licensed under a Creative Commons Attribution 4.0